The objective of this project is to improve the accuracy of measurement of
the proportions and symmetry of polished diamonds.
The commercially available systems for measuring the proportions of polished stones all use a
projection of the profile (shadow method) to scan the diamond. The diamond is rotated between
a telecentric light source and a telecentric lens. The contour of the shadow of the diamond is
then analyzed in a computer to determine its proportions.
This method has some typical disadvantages concerning accuracy.
The ultimate objective of this project was to build a system with much higher accuracy than the
existing systems by using other methods different from the shadow method.
To illustrate the accuracy we use a real life example in combination
with the HRD Antwerp H&A viewer. This perfectly aligned equipment together with
the directional light setup is ideal to compare commercial scanners and
Reflex. The data extracted from the measurement is used to build the
3D-model. These models are visualized with the DiamCalc software from
Octonus. So we have the real image, seen trough the HRD Antwerp H&A viewer, and
the virtual image of the measured diamond, seen in a virtual environment based on the parameters
of the H&A viewer. The creation of H&A patterns is extremely sensitive
to angle variations, and , by nature, very useful as a tool to visualize
slight differences in measurements. Let's start with the arrow side:

The real image shows two incomplete arrows. This is caused by an
internal defectin the crystal lattice, called a glets. This is not
visible in the simulated images because they are based on the external
characteristics only. The center of the images is enlarged to visualize
the differences between the measurements. It is clear that the model
build with Reflex data is far more accurate than the one build from the
data of the commercial scanner. The heart side reveals following pictures:

The blue and green markers
show some differences in the hart and surrounding patterns. The middle
is enlarged to show the importance of a high accuracy measurement.
To test the repeatability we measured the angles of 8 facets of a brilliant 10
times. Between each measurement we removed the diamond and put it
back in a random position. The commercial scanner gave following results:

The results for the Reflex prototype are shown in the next picture:
For less time critical applications like calibration the device
is already suitable. Also the possibility for a future as "proportions
referee" is now being investigated.
Experiments have shown that higher accuracies are possible. Angles (both inclination and index) can be measured with accuracies up to ±0.02º.
The distances can be measured with submicron accuracy. A prototype is being evaluated.